The network module isolator 10/100/1000 MBit/s for the installation in network outlets with RJ45 FM global outlet 2x1 port. The cable whip exits from the isolator back through openings of the cup into the cable channel and connects there with the field jack mounted on the installed network cable. This connection outside of the outlet cup offers the advantage of a simple assembly and while opened the trunk it allows measuring the build-in installation cable with standard This is necessary, since the galvanic separator of the network insulator prevents measuring the whole link from the outlets side. If screened cables are used, the screen can be put to the field-mountable RJ45-socket on the shielding. Field configurable RJ45-jacks exhibits hereunto a the connection with a lug. The network isolation module not continue the screen, since thereby the isolation effect would be waived. module isolator 10/100/1000 MBit/s is conceived on in network outlets with RJ45 modules, e.g. . The cable whip exits from the openings of the cup into the cable channel and connects there with the field-configurable RJ45- on the installed network cable. This connection outside of the outlet cup offers the advantage of a simple assembly and while opened the trunk it allows measuring the in installation cable with standard measuring instruments. ry, since the galvanic separator of the network measuring the whole link from the outlets side. If screened cables are used, the screen can be put to the socket on the shielding. Field- s hereunto a grounding latch for the connection with a lug. The network isolation module does the screen, since thereby the isolation effect.
Technical Data:
| Article Number |
|
2005761 |
| Norms / Standards |
|
IEC 60601-1 resp. EN60601-1:X2006 |
| Electrical strengh between the network connections |
|
4KV |
| connections |
|
1x jack RJ45, 8-pole and 1x plug RJ45, 8-pole |
| Attenuation |
10Mbit/s (0-30MHz) |
-0,8db |
| |
100Mbit/s (0-31,25MHz) |
-0,8db |
| |
1000Mbit/s (0-62,5MHz) |
-2db |
| 1 Network specification |
|
IEEE 803.2 10/100Base-T IEEE 802.3ab 1000Base-T Twisted Pair, auto-conf |
| 1If the link length is > 80m, the transmission power for 1Gbit/s cannot be guaranteed in general. Please check installation first. |
|
|
| Mode of operation |
|
Permanent operation |
| Site conditions during operation |
Air pressure |
860hPa to 1060HPa |
| |
Temperature |
-25°C to 85°C |
| |
Relative humidity |
10% bis 90% (not condensing) |
| Dimensions without cable |
|
(52x23x17)mm³ |
| Cable length |
|
(Standard) 120mm |
| classificaton |
Safety class |
n.a. |
| |
Protection against penetration of water |
No Protection |
| |
Degree of protection of the application part against electrical impact |
n.a. |
| |
Degree of protection against discharge of defibrillators |
n.a. |
| |
Protection for use at presence of inflammable mixtures of anesthetics with air or with oxygen or laughing gas |
No Protection
|
More information, mounting and accessories (PDF File, 600kb)