Produktbeschreibung

Safety: DIN EN 60601-1 (Medical electrical equipment)

EMC: DIN EN 60601-1-2 (Medical electrical equipment)


The MPU100 series of AC/DC switching mode power supplies provide 100 Watts of continuous output power . All supplies are UL94V-1 min compliant, include IEC-320-C14 input for worldwide applications. All models meet FCC Part-18 class B and CISPR-11 EN55011 class B emission Limits and are designed  to comply with UL/c-UL(UL 60601-1) ,TUV/T-mark (EN 60601-1) and new CE requirements. All units are 100% burned in and tested. They are suited for use in hospital instrument and many other applications.

Features:

  • Wide Input Voltage 90 to 264 VAC,47 to 63 Hz
  • IEC-320-C14 Input Inlet
  • Output plug: 2.1mm (inside) x 5.5mm (outside) x 9.5mm (length) barrel plug, straight
  • Output Voltage 19VDC
  • Input Surge Current, Over Voltage And Over Load protection
  • Output Voltage Protection
  • Ultra-low leakage current (under 0.1mA)
  • Power Factor Correction
  • Class I Insulation
  • ON/OFF SWITCH

Electrical Characteristics:

  • Efficiency: 88% max
  • Adj. Output Voltage 16 - 21 VDC
  • Max. Output Current: 6.3 - 4.8 A
  • Total Regulation: 5%
  • Max. Output Power: 100W
  • Line Regulation: 1% max.
  • Load Regulation: 5% max.
  • Hold-up Time: 16mS Typ.
  • Output ripple and noise: 1% (Typ.)

Environmental:

  • Operating Temperature: 0 to 70°C
  • Storage Temperature: -40 to 85°C
  • Relative Humidity: 5 to 95%

    Sym Parameter Test  Conditions Min Typ Max Unit
    Vps Dielectric Withstanding Voltage for Primary  to secondary Primary to secondary 5656     VDC
    Vpg

    Dielectric Withstanding Voltage for Primary  to secondary

    Primary to ground 2800     VDC
    Ris Isolation  Resistance   Io=Full load,Vin=230VAC    50    

    MΩ

    CISPR

    EMI  requirements for CISPR-11   

    Vin=220VAC          

    B     CLASS
    FCC EMI  requirements for  FCC PART-18

    Vin=110VAC

    B     CLASS

     

     

     

     


nach oben