Stromversorgung MED Endo Kit 2

Stromversorgung MED Endo Kit 2
Stromversorgung MED Endo Kit 2
Product No.: 2006390
Please login for prices

Shipping time: 3-4 Days
Print product data sheet 

Produktbeschreibung

Electrical equipment that is used in the patient’s environment or with applied parts, vary according to leakage current relating to the permissible norms by predefined limits (<100μA in normal operation). Under special condition these values can exceed those limits when they are used in a medical electrical system (MES) according to EN 60601-1-1. Device types of classes BF and CF are normally supporting the necessary, but several changes to the power supply of the systems have to be made when those devices are combined with electrical equipment or systems outside the medical used area or patient’s environment.

The medical isolating transformers provide extra protection for patients, users and devices connected to each other. As a leakage current limitation the fulfillment of the requirements for the maximum earth leakage current (normal operation NC <100μA and single fault condition SFC <500μA) becomes possible using the isolation transformers. All devices are also usable for the operation outside medical facilities and rooms.

The devices are in compliance with the following European standards and guidelines: EN60601-1, EN60601-1-1, EN60601-1-2 (EMC) EN14971, MDD 93/42/EEC, Annex I, VII

Scope of supply: 
 

1 x Isolation transformer MED N 2000VA

  • Powder coated metal case color RAL7035
  • Designed for 24/7 use
  • Protection classI, IP21, 
  • Medical. class I, Typ B 
  • Potential equalization according to DIN42801 
  • Secondary output: 9-outlets IEC 320 C13 10A / 250VAC 
  • Max. power: 6,87A 
  • Output 2000 VA 
  • Temperature sensor 
  • switch on current limiter NTC 
  • dimenstions (mm) 385 x 135 x 332 
  • Approvals: EN60601-1,  EN60601-1-1,  EN60601-1-2 , EN14971,  EN980 | MDD 93/42/EWG Anhang I, VII

 
1 x Isolation guard:

Monitoring- and alarm unit for secondary isolation resistance. Optical and acoustical alarm function. Transformator temperature monitoring with optical and acoustical alarm


6 x Med. output power cord 1,8m

Diameter 1mm², length 1,8m, IEC320 C14 straigth to IEC320 C13 straight. Transparent plugs


1 x Input Power cord 5m Orange

Diameter 1mm², length 5m, Schuko plug (international plugs on request) to IEC 320 C13 abgewinkelt


2 x Adapter IEC320 C14 to Schuko

Diameter 1mm², length 1,5m, IEC320 C14 angled, Schuko socket (International sockets on request)


1 x ZPA connector 6mm² / 5m

Lead for additional potential equalization according to DIN 42801. Length: 5m, Diameter: 6mm², lead diameter: 3.8mm, Isolation green-yellow

 

 