New Products

NDS ScaleOR, VGA Input, w/o fiber 90T0013
NDS ScaleOR, VGA Input, w/o fiber 90T0013
Please login for prices
excl. 19 % Tax

Welcome back !

Netzwerk Geräteeinbau Isolator 1000Y

Netzwerk Geräteeinbau Isolator 1000Y
Product No.: 2006158
Please login for prices

Shipping time: 3-4 Days
Print product data sheet 

Produktbeschreibung

The Network Unit Installation Isolator 1000Y provides a safe plug-in of medical devices and personal computers within the patient's vicinity into an IEEE 803.2 (10/100- BaseT,Twisted-Pair) computerNetwork. Workstations are allowed to operate within the patient's, only if they are designed in compliance with IEC 60601-1 as a medical device or if they are fitted with an IEC 60601-1 compliant electrical protection system.


All Inputs and Outputs of a medical device to other non medical devices or Networks outside the surgery environment must comply with IEC-60601-1-1 standard. This galvanic isolation with sufficiently high dielectric strength is required to meet the IEC-60601-1-1 standard. The Network Unit installation Isolator 1000Y may also be vicinity used in non medical applications. It provides galvanic isolation for any equipment connected via Ethernet ports to prevent ground loops.

The Network Unit installation Isolator 1000Y is designed  for mounting in housings with a wall  thickness of up  to 2.5mm. We  recommend connecting  the Faston connector  (6.3x0.8) with the chassis ground, in order to reduce the influence of electric  radiation emitted  from  internal cables.

The Network Unit installation Isolator 1000Y is designed to be used with the Yamaichi connector Y-ConIP20Lock-10, which  provides  improved protection against accidental human contact, and a vibration-stable strain relief. Alternatively, standard RJ45 connectors can be used.

Installation:


The Network Unit Installation Isolator 1000Y can be mounted  into a panel cutout as detailed below, using  the  two mounting screws provided. Optional bezels may be fitted to improve asthetics, or to also improve user and patient safety  in accordance with  IEC 60601-1.
Recommended  tightening  torque: 0.55 Nm.


Technical specification:

Operating mode   non-stop operation
Operating conditions Temperature 1°C to 70°C
  Relative humidity 10% to 90% (non-condensing)
  Air pressure 860hPa to 1060hPa
Storage and Transportation Temperature -25°C to 75°C
  Relative humidity 10% to 90% (non-condensing)
  Air pressure 600hPa to 160hPa
Protection category   IP20
Marking   CE, UL (E249126) RoHS
Network connection   RJ45 socket suitable for use with
Yamaichi Y-CONJACK
Internal Device connections   Horizontal RJ45 socket,
and a 6.3x0.8 Faston tab for connection
to cable shielding
Suitable for use in network environments according to   IEEE 802.3 Cl.14 (10-BaseT),
IEEE 802.3 Cl.25 (100-BaseTX),
IEEE 802.3 Cl.40 (1000-BaseT)
Signal Attenuation 1MHz < f < 62.5MHz <-2db
Rating Insulation voltage   4000V AC
Weight   35g

The Network Unit Installation Isolator 1000Y meets the requirements of the IEC 60601-1 for separators, with 2 Means of Patient Protection (2MOPP)